The effect of a standardised protocol for CIED insertion on complications and infection rates in a DGH

Br J Cardiol 2021;28:73–6doi:10.5837/bjc.2021.027 Leave a comment
Click any image to enlarge
Authors:
First published online 2nd June 2021

Cardiac implantable electronic device (CIED)-related complications and infections typically lead to prolonged hospital stays and, very occasionally, death. A new CIED insertion protocol was implemented in a district general hospital. The primary objective of this study was to determine whether a significant reduction in complication and infection rates occurred after implementation of the new protocol. Medical records were reviewed for patients who had a CIED inserted in the two years pre- and post-protocol implementation, and any complications were identified in a one-year follow-up period.

An increase in the complexity of the devices implanted after introduction of the protocol was observed. The number of complications was significantly reduced from 6.86% to 3.95% (p<0.0001). In the two years prior to protocol implementation, 14 of 871 (1.6%) patients suffered a CIED-related infection. In contrast, four of 683 (0.44%) patients suffered a CIED-related infection in the two years post-implementation. This was not statistically significant (p=0.093).

In conclusion, implementing a standardised protocol for CIED insertion significantly reduced the rate of complications, and also reduced the rate of infection, but this was not statistically significant.

Introduction

There has been a steady increase in the number of cardiac implantable electronic devices (CIEDs) fitted in the UK over the last decade, and common complications of CIED insertion include bleeding or bruising at the insertion site, pneumothorax, lead displacement and infection.

Complications of device insertion are common. Some of the largest datasets come from Denmark. In 2013, an analysis of 5,918 consecutive cases found that 2.4% of patients required a lead-related re-intervention, 0.9% developed a pneumothorax, 0.8% developed infection, 0.6% had a cardiac perforation and 0.2% developed a haematoma requiring re-operation.1 More recently, they have released data on infection in 97,750 consecutive patients and found a rate of infection of 0.82% in the first year after a procedure.2

Infection is a serious complication, which can result in death and prolonged hospital stays.3 There is concern that, due to the increasing complexity of device procedures and the increasing frailty of patients, infection rates are rising disproportionately.2,4,5

This increase in infection is a problem both for the health of the patient and the additional costs to the National Health Service (NHS). Research has shown that implementation of a comprehensive infection control programme can be effective in reducing infection rate.6,7

After an adverse outcome from a patient with a device-related infection at our centre, and a recognition that infection rates seemed unacceptably high, a new, comprehensive CIED standard operating procedure was constructed, with input from multiple departments within the trust, including infection control and microbiology, to try to reduce the number of CIED-related infections in our department. We hypothesised that these changes would reduce the rate of infection.

Method

Table 1. Summary of the new standard operating procedure

All staff to wear hats and masks (inconsistent use before)
MRSA/MSSA screening 5 days before procedure date (new)
Decolonisation with hibiscrub and bactroban for those who were MRSA/MSSA positive (new)
Anaesthetic room/catheter lab locked to minimise the movement of personnel (new)
Chlorhexidine-alcohol was used rather than povidone-iodine for surgical site antisepsis
Patients who had antibiotics did not undergo CIED implantation within 30 days of taking antibiotics, unless there was a pressing clinical indication (new)
Minimum of 18–20 air changes per hour in catheter lab (modifications made to air flow)
Washing of lead aprons and equipment before each procedure (new)
Switch to teicoplanin 6 mg/kg iv, rounded to the nearest 100 mg, + gentamicin 3 mg/kg iv from flucloxacillin 1 g iv + gentamicin 3 mg/kg iv as antibiotic prophylaxis, within 1 hour of the start of the procedure
Increased consultant supervision of procedures and the number of consultants who implanted devices reduced
New drapes purchased, which completely covered the head
Patients with open wounds were not to be operated on until this was resolved, if possible, unless there was a pressing clinical indication (new)
Electrical clippers with single-use head rather than disposable razors used on the day of surgery
Implementation of surgical safety checklist (new)
Patients oxygen saturation was maintained above 95% (new)
Instruments to be inspected prior to use (new)
Key: CIED = cardiac implantable electronic device; iv = intravenous; MRSA = methicillin-resistant Staphylococcus aureus; MSSA = methicillin-susceptible Staphylococcus aureus

The CIED standard operating procedure was implemented in August 2011 (table 1). A comprehensive list of patients who had undergone a CIED procedure within the cardiology department from August 2009 to July 2011 and August 2011 to July 2013 at Musgrove Park Hospital was collated. One year of follow-up after CIED insertion was chosen, as any infection after this point is not considered related to the CIED-insertion procedure.

All patients who underwent a procedure involving a ‘simple’ (single-chamber ventricular pacemaker [VVI], dual-chamber pacemaker [DDD], implantable-loop recorders [ILR]) or ‘complex’ pacing device (implantable cardioverter defibrillator [ICD], cardiac resynchronisation therapy with defibrillator [CRT-D] device, cardiac resynchronisation therapy with pacemaker [CRT-P]) were included. Whether or not it was a new procedure, a box change, a revision of an existing device or a procedure related to a complication was recorded. Data were acquired from yearly audit spreadsheets and an electronic implant database, which were then compared with the paper diaries kept by the catheter lab to ensure all procedures were documented. Operator details were also recorded where available.

Each patient’s paper and electronic medical records, including X-rays and discharge summaries, were reviewed to determine whether the patient had experienced any complications from their procedure. A complication was defined as any adverse event recorded within 12 months of the CIED insertion. The following complications were recorded:

  • Infection requiring antibiotics and device removal or palliative care if not considered fit for device removal; superficial skin infections were not included
  • Pneumothorax, whether or not a drain was required
  • Haematoma requiring re-operation
  • Lead displacement, whether or not requiring lead repositioning
  • New pericardial effusion, whether or not requiring drain
  • A ‘failed’ procedure – failure to implant a left ventricular (LV) lead
  • Other significant complications – detailed.

The percentages of infection and complication rates were calculated for each group. Fisher’s exact test was used to test for significance. The relative risk was calculated to determine the risk of a patient from each group developing an infection and confidence intervals (CI) were then calculated.

Results

A total of 1,554 procedures took place over the four-year period. In 2009 to 2011 there were 871 procedures: 520 (59.7%) were male; 721 (82.8%) procedures involved ‘simple’ devices and 150 ‘complex’. A more complete breakdown of the procedures is shown in table 2. In 2011 to 2013 there were fewer procedures – 683: 429 (62.8%) were male; 506 (74.1%) procedures involved ‘simple’ devices and 177 ‘complex’. This increase in the proportion of complex devices was highly significant (p<0.0001).

Table 2. Procedures undertaken

2009 to 2011 2011 to 2013
Procedures % Procedures %
New 601 69.0 500 73.2
Revision 38 4.4 43 6.3
Complication* 38 4.4 18 2.6
Box change 193 22.2 122 17.9
Explant 1 0.1 0 0.0
Total 871 100 683 100
Device type
ILR 114 13.1 74 10.8
AAIR 1 0 0.0
VVIR 158 18.1 129 18.9
DDDR 448 51.4 303 44.4
CRT-P 29 3.3 31 4.5
VVI ICD 23 2.6 28 4.1
DDD ICD 63 7.2 63 9.2
CRT-D 34 3.9 54 7.9
Subcutaneous ICD 1 0.1 1 0.1
Total 871 100 683 100
Procedure type
ILR 114 13.1 74 10.8
‘Simple’ 607 69.7 432 63.3
‘Simple’/ILR 721 82.8 506 74.1
‘Complex’ 150 17.2 177 25.9
Total 871 100 683 100
*Complication – refers to a procedure to rectify a complication that has occurred.
Key:AAIR = single-chamber atrial pacemaker; CRT = cardiac resynchronisation therapy; D = defibrillator; DDD(R) = dual-chamber pacemaker; ICD = implantable cardioverter-defibrillator; ILR = implantable loop recorder; P = pacemaker; VVI(R) = single-chamber ventricular pacemaker

There were 15 different operators in the pre-changes period, six of whom were consultants. There were nine different operators in the post-changes period, four of whom were consultants (table 3). Note that there was a small percentage (3%) of unknown operators in the 2009 to 2011 period due to missing data.

In the pre-changes group, 60 (6.9%) patients experienced a complication and 38 required a procedure to rectify a complication (4.4%). In the post-changes group, 27 (4.0%) experienced a complication and 18 required a procedure to rectify a complication (2.6%) (table 4). In both time periods, atrial lead displacement was the most common complication. In the pre-changes group, 14 (1.6%) patients had a CIED-related infection, which resulted in surgical removal of the device. In the post-changes group, there were four (0.6%) patients who had a CIED-related infection, which resulted in surgical removal of the device. This difference was not quite statistically significant (p=0.093).

Discussion

We introduced a standard operating procedure, including a surgical safety checklist, with the principal aim to reduce CIED-related infection. A reduction in infection rates was observed, but this was not quite statistically significant. Previous studies have found that implementation of a protocol to reduce CIED-related infection is effective.9,12 This audit corroborates these earlier findings and provides further evidence that there is potential for national guidelines to have an effect on CIED-related infection rates. However, complication rates were reduced significantly; the overall rate of complications was reduced by over 40%. Surgical safety checklists have been shown to reduce complications.8

Prevention of complications is important at a time when the number of CIEDs being implanted is increasing. It impacts on patients, as many need further procedures, which is inefficient. Early re-intervention also increases infection risk.9 We believe that the reduction in complications occurred because of the constrained number of operators, increased surveillance of junior doctors and the introduction of a checklist.

Prior to this audit there were no fixed guidelines for CIED-related infection prevention at Musgrove Park Hospital, and nationally there were no set guidelines/protocols for CIED implantation. After the introduction of the new protocol there was a reduction in patients who contracted a CIED-related infection, this did not quite meet statistical significance. It is unclear what improved outcomes here. Previous research has shown operator experience reduces CIED-infection rates.5 There are little data on the most optimal prophylactic antibiotics, but pre-operative antibiotics appear effective.10,11 The change to chlorhexidine skin preparation was driven by a trial suggesting it was more effective than povidone-iodine.12 National Institute for Health and Care Excellence (NICE) clinical guideline 74 (now updated to NG125) was also referred to in the design of the new protocol.13

CIED-related infection significantly increases both admission and long-term mortality rates.14 This reduction in infection would result in a financial saving to the NHS, with 10 less patients over a two-year period potentially requiring further surgery and prolonged inpatient stays. The average cost of CIED-related infection in the UK is £30,958,7 and, therefore, a reduction of 10 patients with CIED-related infections over the two years would be a potential cost saving of £309,580. This would translate to a potential long-term saving to the NHS Trust of approximately £150,000 per annum.

Limitations and future directions

This research was carried out retrospectively, which, therefore, bears the inherent limitations of such studies, including patients who may have been lost to follow-up and errors in classification or recording. These limitations, and any bias, were minimised by including all procedures carried out within the time period and by using as many avenues as possible to follow-up the patients, including reviewing paper and electronic notes, device checks and X-ray imaging. Retrospective data collection limits the accessible data, so variables such as procedure time, for example, were not available, limiting the analysis of potential confounders.

The protocol implemented multiple changes in practice, therefore, it is not possible to determine which factors led to the reduction in complication rates, and which were most important. However, as the majority of changes implemented were fairly simple and low cost this should not be problematic, and the benefit of implementing the protocol would likely outweigh this disadvantage.

More recently, antibiotic envelopes have become available. The WRAP-IT (Worldwide Randomized Antibiotic Envelope Infection Prevention Trial) study showed that for complex procedures, an antibiotic envelope could reduce the rate of infection (from 1.2% to 0.7%, p=0.04).15 Interestingly, just one more infection in the envelope group would have rendered the results non-significant (p=0.052). Whether this is cost-effective in the UK remains to be seen. In the US the base-case incremental cost-effectiveness ratio of the antibacterial envelope compared with standard-of-care was $112,603 per quality-adjusted life year,16 above the threshold typically set by NICE.

Finally, as this research was undertaken in a single-centre district general hospital, generalising conclusions from these findings across all centres would not be appropriate. We hope that a national audit, as takes place in some countries, would give a better understanding of the national infection rates for CIEDs, and that adoption and iteration of guidelines will continue to reduce complication rates.17

Conclusion

In summary, a comprehensive change in practice with the aim to reduce infection rates was successful, and also had the added and unforeseen benefit of reducing overall complication rates from device insertion.

Key messages

  • Cardiac implantable electronic device (CIED) insertion results in significant complications in a small proportion of individuals, causing morbidity, cost to the NHS and occasionally mortality
  • A comprehensive change in clinical practice significantly reduced CIED-related complications
  • A reduction in infection rates was seen, which was the principal aim of the new guidance, but this did not reach statistical significance

Conflicts of interest

MD reports receiving speaker fees from Biotronik in the past 2 years. EF, GF: none declared.

Funding

None.

Study approval

All research adhered to the World Medical Association’s declaration of Helsinki. No national ethical approvals were required and local approvals were secured.

References

1. Kirkfeldt RE, Johansen JB, Nohr EA, Jorgensen OD, Nielsen JC. Complications after cardiac implantable electronic device implantations: an analysis of a complete, nationwide cohort in Denmark. Eur Heart J 2014;35:1186–94. https://doi.org/10.1093/eurheartj/eht511

2. Olsen T, Jorgensen OD, Nielsen JC, Thogersen AM, Philbert BT, Johansen JB. Incidence of device-related infection in 97 750 patients: clinical data from the complete Danish device-cohort (1982–2018). Eur Heart J 2019;40:1862–9. https://doi.org/10.1093/eurheartj/ehz316

3. Greenspon AJ, Patel JD, Lau E et al. 16-year trends in the infection burden for pacemakers and implantable cardioverter-defibrillators in the United States 1993 to 2008. J Am Coll Cardiol 2011;58:1001–06. https://doi.org/10.1016/j.jacc.2011.04.033

4. Tarakji KG, Ellis CR, Defaye P, Kennergren C. Cardiac implantable electronic device infection in patients at risk. Arrhythm Electrophysiol Rev 2016;5:65–71. https://doi.org/10.15420/aer.2015.27.2

5. Dai M, Cai C, Vaibhav V et al. Trends of cardiovascular implantable electronic device infection in 3 decades: a population-based study. JACC Clin Electrophysiol 2019;5:1071–80. https://doi.org/10.1016/j.jacep.2019.06.016

6. Borer A, Gilad J, Hyam E et al. Prevention of infections associated with permanent cardiac antiarrhythmic devices by implementation of a comprehensive infection control program. Infect Control Hosp Epidemiol 2004;25:492–7. https://doi.org/10.1086/502428

7. Ahsan SY, Saberwal B, Lambiase PD et al. A simple infection-control protocol to reduce serious cardiac device infections. Europace 2014;16:1482–9. https://doi.org/10.1093/europace/euu126

8. Haynes AB, Weiser TG, Berry WR et al. A surgical safety checklist to reduce morbidity and mortality in a global population. N Engl J Med 2009;360:491–9. https://doi.org/10.1056/NEJMsa0810119

9. Klug D, Balde M, Pavin D et al. Risk factors related to infections of implanted pacemakers and cardioverter-defibrillators: results of a large prospective study. Circulation 2007;116:1349–55. https://doi.org/10.1161/CIRCULATIONAHA.106.678664

10. de Oliveira JC, Martinelli M, Nishioka SA et al. Efficacy of antibiotic prophylaxis before the implantation of pacemakers and cardioverter-defibrillators: results of a large, prospective, randomized, double-blinded, placebo-controlled trial. Circ Arrhythm Electrophysiol 2009;2:29–34. https://doi.org/10.1161/CIRCEP.108.795906

11. Da Costa A, Kirkorian G, Cucherat M et al. Antibiotic prophylaxis for permanent pacemaker implantation: a meta-analysis. Circulation 1998;97:1796–801. https://doi.org/10.1161/01.CIR.97.18.1796

12. Darouiche RO, Wall MJ Jr., Itani KM et al. Chlorhexidine-alcohol versus povidone-iodine for surgical-site antisepsis. N Engl J Med 2010;362:18–26. https://doi.org/10.1056/NEJMoa0810988

13. National Institute for Health and Care Excellence. Surgical site infections: prevention and treatment. London: NICE, 2008. Updated guidance available from: https://www.nice.org.uk/guidance/ng125

14. Sohail MR, Henrikson CA, Braid-Forbes MJ, Forbes KF, Lerner DJ. Mortality and cost associated with cardiovascular implantable electronic device infections. Arch Intern Med 2011;171:1821–8. https://doi.org/10.1001/archinternmed.2011.441

15. Tarakji KG, Mittal S, Kennergren C et al. Antibacterial envelope to prevent cardiac implantable device infection. N Engl J Med 2019;380:1895–905. https://doi.org/10.1056/NEJMoa1901111

16. Wilkoff BL, Boriani G, Mittal S et al. Cost-effectiveness of an antibacterial envelope for cardiac implantable electronic device infection prevention in the US healthcare system from the WRAP-IT trial. Circ Arrhythm Electrophysiol 2020;13:e008503. https://doi.org/10.1161/CIRCEP.120.008503

17. Sandoe JA, Barlow G, Chambers JB et al. Guidelines for the diagnosis, prevention and management of implantable cardiac electronic device infection. Report of a joint Working Party project on behalf of the British Society for Antimicrobial Chemotherapy (BSAC, host organization), British Heart Rhythm Society (BHRS), British Cardiovascular Society (BCS), British Heart Valve Society (BHVS) and British Society for Echocardiography (BSE). J Antimicrob Chemother 2015;70:325–59. https://doi.org/10.1093/jac/dku383

THERE ARE CURRENTLY NO COMMENTS FOR THIS ARTICLE - LEAVE A COMMENT